The Randox Adhesion Molecule array allows for multiplex analysis of E, L and P Selectin, ICAM-I and VCAM-I. Allowing researchers interested in the effects of numerous adhesion molecules to receive simultaneous results in real time.

Adhesion Molecules

Analyte Intra-Assay Precision (n=20) Intra-Assay Precision (n=20)
  Level 1 %CV Level 2 %CV Level 3 %CV Level 1 %CV Level 2 %CV Level 3 %CV
E-Selectin 7.0 7.4 5.2 7.5 8.0 8.6
L-Selectin 6.2 6.1 9.1 8.7 7.7 13.4
P-Selectin 6.5 6.1 9.6 5.7 4.8 7.2
ICAM-I 6.4 8.4 9.4 3.5 7.9 8.3
VCAM-I 7.7 8.1 9.2 5.9 9.6 7.5
Analyte Calibration Range ng/ml Sensitivity ng/ml
E-Selectin 0-25 0.1
L-Selectin 0-350 3.2
P-Selectin 0-120 1.9
ICAM-I 0-100 1.7
VCAM-I 0-330 4.1

Accuracy and Precision based on Evidence Investigator

Assay range may vary with batch of calibrators.
Inter-assay precision was determined by assaying 2 replicates of each of three levels of sample in 10 separate assays.
Intra-assay precision was determined by assaying 20 replicates of each of three levels of sample.